HVP-40M Digital High voltage meter Product ID: HVP-40M
Model: HVP-40DM ( 40KVDC, 2000MΩ , 3999LCD DMM )
Features:
DC 40KV CAT II Pollution 2.
2000MΩ Impedance.
3999 LCD display.
Accuracy 1.0% full range.
T.C. 100 ppm/oC
400V/4KV/40KV, 3 range.
Resolution 0.1V/1V/10V.
PEAK HOLD, Auto power off.
CE, IEC1010
HVP-40M Analog High Voltag probe
- SPECIFICATIONS ( 20 ± 5°C, RH < 80% )
- ! WARNING !
- Before taking any measurements, first connect the alligator clip of this probe to earth ground and make sure connection is electrically good.
- Input impedance:600M Ohm
- Maximum operation voltage:40KV DC , CAT II , Pollution 2
- Polarity:Positive only
- Display:Analog indicate
- Accuracy:Factory calibrater 25KV ± 2%
- FULL scale 40KV ± 3 %
- Temperature Coefficient:Maximum loading Current:Maximum loading power:Voltage Range:40KV / 1 Range
- Voltage resolution:1KV
- Power source:Need not power
- Operating temperature:0 °C ~ + 50 °C.
- Storage temperature:- 20 °C ~ + 70 °C.
- Ground lead length:90cm ( 35 " )
- Dimensions:420 L x 80 ø
- Weight:360g
Product ID: Tecpel HVP-40
Features:
DC 40KV CAT II Pollution 2
600M Ohm Impedance
25KV ± 2%, 40KV ± 3%
T.C. 200 ppm/°C.
Positive Polarity only.
Resolution 1KV
No Need power.
CE, IEC1010