CL-327A Process Calibrator Product ID: CL-327A
The NEW Tecpel® CL-327A is a battery-powered, hand-held process calibrator that allows process technicians to simulate and source variables signals. Output signals including DC voltage, DC current, ohms, analog transmitters, thermocouple, RTD, frequency, pulse, Switch On-OFF, are widely used in measurement and calibration for industrial field and laboratory application, it is one of necessary tools used to test and calibrate industrial automation systems and instruments. Get your job done fast and get it done precisely at affordable price. Your right choice!
Process Calibrator CL-327A
general characteristics:
1, The basic accuracy of the output of up to 0.02%, the output is displayed as 6 digit
2, The output functions: DC voltage, DC current, ohms, analog transmitters, thermocouple, RTD, frequency, pulse, Switch On-OFF
3, The DC current output, can provide 25 percent and 100 percent of manual stepping, automatic step and auto ramp output function
4, The thermocouple output, can provide high-precision automatic cold junction compensation, ℃ or ℉ temperature display
5, Outside with high-precision temperature probe, accuracy: ± 0.2 ℃
6, Large-screen LCD display multiple data can be displayed simultaneously: DC current and % values, thermocouples, temperature values of thermal resistance and the corresponding indexing values, etc.
7, The operating performance is superior keyboard configuration, set the value corresponding to the output bit setting change keys and LCD display.
*Working temperature and humidity
0 to 50ºC, ≤80% RH non-condensing
40 to 50ºC, ≤70% RH
*Storage temperature and humidity
-25 To 60ºC, ≤90% RH non-condensing
*Display and backlight
* Segment LCD dual display
* White LED backlight
*power supply
* 4 x 1.5V AAA alkaline batteries
* Automatic shutdown: can set auto-off time from 0 to 60 minutes
*Size and Weight
205 × 95 × 42 (mm), approximately 500g
Features; High Accuracy of 0.02% for source, 6 digits display for source.A dual,
Liquid Crystal Display;White LED for backlight
Function |
Reference |
Range |
Resolution |
Accuracy |
Remark |
Thermocouple |
R |
0℃~1767℃ |
1℃ |
0~100℃:1.5℃ |
By using ITS-90 temperature scale;The accuracy does not include the error of internal temperature compensation caused by a sensor; |
100~1767℃:1.2℃ |
|||||
S |
0℃~1767℃ |
1℃ |
0~100℃:1.5℃ |
||
100~1767℃:1.2℃ |
|||||
B |
600℃~1820℃ |
1℃ |
600~800℃:1.5℃ |
||
800~1820℃:1.1℃ |
|||||
E |
-200.0℃~1000.0℃ |
0.1℃ |
-200.0~-100.0℃:0.6℃ |
||
-100.0~600.0℃: 0.5℃ |
|||||
600.0~1000.0℃: 0.4℃ |
|||||
K |
-200.0℃~1372.0℃ |
0.1℃ |
-200.0~-100.0℃: 0.6℃ |
||
-100.0~400.0℃: 0.5℃ |
|||||
400.0~1200.0℃: 0.7℃ |
|||||
1200.0~1372.0℃: 0.9℃ |
|||||
J |
-200.0℃~1200.0℃ |
0.1℃ |
-200.0~-100.0℃: 0.6℃ |
||
-100.0~800.0℃: 0.5℃ |
|||||
800.0~1200.0℃: 0.7℃ |
|||||
T |
-250.0℃~400.0℃ |
0.1℃ |
-250.0~400.0℃:0.6℃ |
||
N |
-200.0℃~1300.0℃ |
0.1℃ |
-200.0~-100.0℃: 1.0℃ |
||
-100.0~900.0℃: 0.7℃ |
|||||
900.0~1300.0℃: 0.8℃ |
|||||
L |
-200.0℃~900.0℃ |
0.1℃ |
-200.0~0.0℃:0.7℃ |
||
0.0~900.0℃:0.5℃ |
|||||
U |
-200.0℃~600.0℃ |
0.1℃ |
-200.0~0.0℃:0.7℃ |
||
0.0~600.0℃: 0.5℃ |
Other feature:
1、Temperature Coefficient: 0.1 times the applicable accuracy,specification per degree
C for 5℃to 18℃and 28℃to 50℃.
2、The range of the internal temperature compensation sensor is
from 0℃to 40℃, Compensation error ≤±0.5℃.
3、The accuracy of the temperature probe: ±0.2℃.
The range of the measured temperature is form -20℃-100℃.
4、Maximum voltage between any output terminal and earth: 30Vp-p
5、Maximum output current: Approximately 25mA.