Laboratory DC Power Supply(GPS-3030D/ GPS-3030DD) Product ID: GPS-3030D/ GPS-3030DD
GPS-SERIES
(DIGITAL TYPE)
GPS-3030DD (DIGITAL TYPE)
1. Light and Compact Design
2.Remote Control by External Voltage
3. 0.01% High Regulation
4.Internal Select for Continuous or Dynamic Load
5.Low Ripple and Noise
6. Series or Parallel Operation Function Overload and Reverse Polarity Protection
- SPECIFICATIONS:
- CONSTANT VOLTAGE
- OPERATION Regulation Line regulation < 0.01% + 3mV
- Load regulation < 0.01% + 3mV (rating current < 3A)
- < 0.02% + 5mV (rating current > 3A)
- Ripple & Noise < 0.5mVrms 5Hz ~ 1MHz (rating current < 3A)
- < 1mVrms 5Hz ~ 1MHz (rating current > 3A)
- Recovery Time < 100µS (50% Load change, Minimum load 0.5A)
- Temp. Coefficient < 300ppm/°C
- Output Range 0 to rating voltage continuously adjustable
- --------------------------------------
- CONSTANT CURRENT
- OPERATION Regulation Line regulation < 0.2% + 3mA
- Load regulation < 0.2% + 3mA
- Ripple Current < 3mArms
- Output Range 0 to rating amperes continuously adjustable
- (Hi / Lo range switchable)
- --------------------------------------
- Digital 3 1/2 digits 0.5"" LED display
- 3 1/2 digits 0.39"" LED display (for GPS-3030DD)
- Accuracy + (0.5% of rdg + 2 digits)
- --------------------------------------
- INSULATION Chassis and Terminal 20MΩ or above ( DC 500V )
- Chassis and AC Cord 30MΩ or above ( DC 500V )
- --------------------------------------
- POWER SOURCE AC 100V/120V/220V/240V+10%, 50/60Hz
- --------------------------------------
- ACCESSORIES Instruction manual x 1, Power cord x 1
- Test lead GTL-105 x 1 (--------------------------------------
- DIMENSIONS 128(W) x 145(H) x 285(D) mm
- SELECTION GUIDE
- Model Output Volts (V) Output Amps (A)
- Digital
- GPS-3030D (Single LCD Display) 0 ~ 30V ;0 ~ 3 A
- GPS-3030DD(Dual LCD Display) 0 ~ 30V; 0 ~ 3 A