K Type Temperature Thermocouples(TPK-04L / TPK-04LS) Produkte ID: TPK-04L / TPK-04LS

Kategorie: Thermoelemente
Zertifikat:
  • TPK-04LS 90 degree Short Angle Surface Probe
    • Measurement range: -50 ° C to 500 ° C.( -58 ° F to 932 ° F)
    • Diameter: 14 mm
    • Length : 185 mm (Stainless steel)
    • Hand Holder: 120 mm
    • Coiled Cable: 1500 mm
    • Short Angle Arm length: 25mm
  • TPK-04L 90 degree Long Angle Surface Probe
    • Measurement range: -50 ° C to 500° C.( -58 ° F to 932 ° F)
    • Diameter: 14 mm
    • Length : 185 mm (Stainless steel)
    • Hand Holder: 120 mm
    • Coiled Cable: 1500 mm
    • Long Angle Arm length : 63mm
  • Difference between TPK-04LS and TPK-04L
    • TPK-04LS :
    • Short Angle Arm length: 25mm
    • TPK-04L:
    • Long Angle Arm length : 63mm

How to use Surface temperature thermocouple correctly?

 

  1. The Objective surface should be kept clean, as a result of oil, dust, or other insulating material. For example,  paint will affect the measurement accuracy .
  2. Measured surface should be flat, For example,  the measured object is rough or uneven, you need to use grease ,smear on its surface.to get better heat transmitting. Rough surface tends to cause the thermocouple sensor damaged or poor performance of  the accuracy .
  3. Measurement of surface area must be greater than the size of a probe sensor.
  4. Prohibit using surface thermocouple for rolling objects. Suggest to use scroll wheel temperature measurements Tecpel TPK-06 Rolling thermocouple to measuring rolling objectives.
  5. Take the vertical angle of 90 ° degrees.  The error should not exceed 5 ° degree angle to ensure the accuracy of temperature measurement
  6. 6. Do not measure the temperature exceeds its rated temperature, to avoid the thermocouple from being damaged.